Unknown operator tag M-920
The system is set up to require use of the automatic identication system for operator identication.
Cause: The read tag is entered incorrectly or is not reported in the management database.
Action: Report the situation to the MEDIVATORS
®
ADVANTAGE PLUS
®
Automated Endoscope
Reprocessor manager along with a request to add the correct tag code to the operator list.
The scanned tag is an unknown endoscope code M-922
The system is set up to require use of the automatic identication system for endoscope selection.
Cause: The read tag is entered incorrectly or is not reported in the management database.
Action: Report the situation to the MEDIVATORS ADVANTAGE PLUS System manager along with a
request to add the correct tag code to the endoscope list.
Barcode not found M-930
Cause: The barcode was read but the OK
button was not enabled.
Action: Click the OK
button.
The code “%s” is not the correct %s code. M-940
Cause: An incorrect code or lot number was entered while performing the replace uids utility for
either the detergent or alcohol.
Action: Re-run the replace uids utility and enter a valid code.
Endoscope Errors
Start not allowed: Vent hookup block S-102
This message appears if no disinfection program is active and the leak test sensor is on. When this message
occurs, no disinfection program can be started.
Cause: If the endoscope is vented too quickly, then it remains pressurized, causing the leak test sensor
to be on.
Action: Stop the program and disengage the terminal block to reduce the pressure. This causes the
message W-751 to disappear. If the error message remains, see the next cause and action below.
Cause: A faulty sensor faulty continues to incorrectly read on.
Action: Report the situation to the maintenance department for repairs.
Cause: The air valve leaks.
Action: Report the error to the maintenance department.