Endoscope leak or connected incorrectly S-103
Cause: The endoscope or connection hose is leaking seriously. The leak test connection is not in order.
Action: Use a manual leak tester (with a pressure gauge) to see if the endoscope leaks. If so, have
the endoscope repaired. If there is no leak, check the endoscope connection and restart the
program. If the report recurs, consult the maintenance department.
Cause: The block is out of alignment to the connection plate in the rinse basin.
Action: Stop the program, disconnect, and place the block anew. Restart the program.
Minor endoscope leak, cycle continuing safely S-104
Cause: Very tiny leak in the endoscope.
Action: Using the pressure gauge, pump the endoscope to pressurize and check to see if it is leaking.
If there is a leak, report it to the maintenance department. If the endoscope is not leaking, see
other causes of the error below.
Cause: Very tiny leak in the leak test system of the MEDIVATORS
®
ADVANTAGE PLUS
®
Automated
Endoscope Reprocessor, the terminal connector, or the connection block.
Action: Consult the maintenance department to repair the leakage.
Cause: A power failure vents the endoscope and the error message follows.
Action: If the machine automatically started up after the error, no further action is required. Otherwise,
contact the maintenance department.
Cause: If the endoscope stays pressurized, a sensor is faulty.
Action: Report the situation to the maintenance department.
Leak test vented by pressure remains high S-105
Cause: There may be an air valve defect if air does not vent from the endoscope during the ventilation
process.
Action: Try to start the process again by pressing the Start
button. If the message recurs, press
the Stop
button. Remove the connection block and check to see if air is escaping (there is
a hissing noise). If so, report it to the maintenance department. If air is not escaping, see other
possible causes of the error below.
Cause: The leak test sensor is defective or out of alignment. This causes the signal to be maintained
after the block is removed.
Action: Report to the maintenance department for repairs.
Cause: The ow control air valve is contaminated or out of alignment.
Action: Try to start the process step again by pressing the Start
button. If the error recurs, ask the
maintenance service to align the valve.