116
Block not pressurized or not connected S-106
Cause: The endoscope is not properly connected.
Action: Stop the machine and check if the leak test connections and the terminal block are properly
placed. Connect the leak test and block again. Start the program.
Cause: Machine fault.
Action: Consult the maintenance department.
Start not allowed: Vent hookup block
S-112, S-122, S-132,
S-142, S-152, S-162,
S-172
This report occurs for the following channels:
• Air • Suction • Elevator
• Biopsy • Jet
• Extra • Water
These errors appear if a disinfection program is selected and the specic channel sensor is on. When this
message occurs no disinfection program can be started.
Stop the program and open the lid. An endoscope warning (W-764 through W-770) appears. This message
must be cleared before a restart takes place.
Cause: The endoscope channel is blocked, causing the sensor to be on.
Action: Disconnect the connector from the channel in the endoscope. If the message disappears, report
the blocked endoscope to the maintenance department. If the message remains, see other
possible causes of the error below.
Cause: The connection block is blocked when the endoscope is not connected.
Action: Disconnect the connection block and replace it with another. If the message disappears, let the
maintenance department unclog the block. If the message remains, see other possible causes of
the error below.
Cause: The channel sensor is defective or out of alignment
Action: Report the faulty sensor to the maintenance department.