77
Taking a Sample of Disinfectant – Check MRC
At the end of a cycle, the system will prompt to collect disinfectant.
1. Open the center door to access the sample port.
2. Wearing protective gloves, take a sample of the disinfectant from the sample port by pressing a sample
cup up against the metal prong.
Figure 28 Taking a Disinfectant Sample
3. Close the access door.
4. Test the sample using RAPICIDE
®
PA Test Strip to ensure that the disinfectant is above MRC.