Fluid System
Fluid Reservoir Status
The system monitors the use of disinfectant, alcohol, and detergent. The percentage of uid remaining
appears in the footer of all screens.
Figure 24 Reservoir Levels
WARNING Relling the container with the incorrect uid may cause the endoscope to be
improperly disinfected. Do not use the endoscope on a patient if this occurs.
Wear protective clothing, gloves and eye wear. The disinfectant and detergent manufacturers
may recommend additional protection.
If the uid reservoir status indicates that the detergent or alcohol volume is low, then you must follow the
alcohol or detergent loading process to ll the appropriate uid reservoir.
Note: A program does not stop if a 0% volume level is reached during a cycle. An error message appears
and you must replace the uids.
The uid levels in the detergent and alcohol reservoirs are monitored with oat switches. A cycle cannot
be initiated if sufcient uid is not available. The replace uid utilities must be run once out of uid is
detected even though the reservoirs have been relled.