O-Ring Lubrication
1. Using a cloth, dry the O-rings on the connection columns. Check for wear and replace as needed.
2. Take a small amount of O-ring grease between your thumb and index nger and spread it diagonally
on the connector plate along the rst two columns.
3. Repeat these two steps for the other three pairs of columns.
4. Spread the grease lengthwise along the O-rings with your thumb and index nger.
Monthly Maintenance
Basin Drain Screen
Using a screen hook, remove and clean the screen. Check for wear and replace as needed.
Air Filter Replacement
The MEDIVATORS
®
ADVANTAGE PLUS
®
Automated Endoscope Reprocessor has four (4) air lters.
Each lter should be replaced every 6 months.
1. Acquire new inline lter assemblies (See Appendix B for replacement part numbers).
2. Open the right and left blue-doors and locate the lters.