76
Status: Shutdown
The endoscope cannot be removed from the machine. Continue the disinfection program.
After the current program is completed, run the disinfection program again. Except for the operator login
name, all entered data are automatically copied from the previous run if these steps are followed:
1. Press the Open/Close
button.
2. Select your login name and press the OK
button.
3. Enter your login name and password.
4. The basin cover opens.
5. Within 30 seconds, press the Menu
button.
6. Select your login name and press the OK
button.
7. Enter your login name and password.
8. Select “Repeat last run” and press the OK button.
System Errors
If a system error occurs during standby mode, a warning message appears as the fault occurs but disappears
as the situation is corrected.
If the error occurs during a disinfection cycle, the program ends and uids are drained from the basin.
A recovery program runs until a safe state is reached and the endoscope can be removed. The recovery
program cannot be stopped by the operator.
WARNING Report all system errors to your supervisor.