Program Errors
Restarting a Program
In a certain instances, a program may be restarted after an error is displayed to restart the program if the
following conditions are true:
To restart the program:
1. Press the Start
button.
2. If the program restarts and progresses without any further errors, the disinfection cycle is valid.
3. If the program does not restart, or one or more errors appear, press the Cancel button. Follow
the instructions below for the steps to take next.
If a program is stopped by an error condition and lasts longer than 60 minutes, the program can not be
restarted, the LED turns red, and the program status is ASC (Reject).
Canceling a Program
If the Cancel
button is pressed, the program runs until a safe state is reached.
The alarm sounds and cancel message appears in the Report section of the screen.
Every time the Cancel button is pressed, the current program step is stopped and the next step in
the cancel sequence is called. By pressing the Cancel button multiple times, the program can be
forced to the end of the cycle. However, caution must be taken because the endoscope channels remain
lled with uid.
One of two status messages appears on the monitor. Follow the instructions below for the status message
on your monitor.
Status: Failure Recovery
1. Press the Cancel button.
2. Press the Open/Close
button.
3. Select your login name and press the OK
button.
4. Enter your login name and password.
5. The basin lid opens.
6. Remove the endoscope or correct the cause of the error.
7. Re-start the disinfection program.