19
WARNING Use only uids approved by Medivators Inc.
Check the expiration date of the disinfectant in the reprocessor each day.
Never use disinfectant beyond the manufacturer’s expiration date, even if the MRC level is
acceptable.
Never use disinfectant that tests below the acceptable MRC level, even if it has not yet reached
the expiration date.
When replacing disinfectant, wear personal protective equipment (clothing, mask, gloves & eye
wear). The disinfectant and detergent manufacturers may recommend additional protection.
If leaking liquid is present and you suspect it is disinfectant, read and follow the disinfectant
instructions before attempting a cleanup.
WARNING Ensure that the reprocessor is installed and maintained in a position that allows
access to the power cord for emergency disconnection if necessary.
Never use the equipment in a manner inconsistent with the manufacturer’s guidelines. Using
the equipment in a manner inconsistent with the manufacturer’s guidelines may impair the
protections that are designed into the device and may cause injury to the user.
RAPICIDE
®
PA containers must be triple rinsed before disposal.