118
Blocked channel
S-114, S-124, S-134,
S-144, S-154, S-164,
S-174
This report occurs for the following channels:
• Air • Suction • Elevator
• Biopsy • Jet
• Extra • Water
Try to start the process again by pressing the Start button. This can be done once. If the message recurs,
the Stop
button must be pressed.
Cause: Connection hose(s) is kinked or pinched.
Action: Check all hoses. If kinked or pinched, straighten out hose(s) and start the program again. If
none of the hoses is kinked or pinched, see other possible causes of the error below.
Cause: Channel blocked.
Action: If the channel is blocked, have the endoscope checked or repaired. If there is no blockage, try
the endoscope in a different MEDIVATORS
®
ADVANTAGE PLUS
®
Automated Endoscope
Reprocessor. If the error is repeated, check the connection block. See other possible causes of
the error below.
Cause: Connection block obstructed.
Action: Press a properly tting hose of a small bellows pump (manual leak test) in the holes on the
underside of the terminal block. Squeeze the bellows. Air should come out of the channel
connection. If there is a blockage, consult the instrument service manual to unclog the
connection block. If the block is not obstructed, then a machine fault is probably the cause.
Consult the maintenance department.
No pressure: channel
S-115, S-125, S-135,
S-145, S-155, S-165,
S175
This report occurs for the following channels:
• Air • Suction • Elevator
• Biopsy • Jet
• Extra • Water
Cause: There is insufcient air pressure to blow the moisture out of the channels.
Action: Press the Start
button to try the process step again. If the message recurs, press the Stop
button.
Cause: A hose is detached.
Action: Verify all hoses are still connected to the endoscope then start the program again.
Cause: Incorrect parameter set.
Action: When in doubt or if the wrong parameter is set, report the problem to the MEDIVATORS