162
MEDIVATORS
®
ADVANTAGE PLUS
®
Automated Endoscope Reprocessor machines are selected by
their serial number. This can be found on the name plate inside the front door. To select a machine and its
parameters, follow these steps:
1. Select the machine by clicking on the serial number in the System Overview window.
2. Click on the Open button on the icon bar.
3. The parameter overview appears in the Parameter Overview window.
4. A description of each parameter function appears in the tables below.
Table 1 System Parameters
Parameter Function
For patient disinfection data call-ups
If set to “Yes,” the requirement to enter patient data is activated
and appears on the screen.
Specialist/assistant
If set to “Yes,” the requirement to enter physician or assistant data
is activated and appears on the screen.
Sort list by name
If set to “Yes,” overviews for physicians and assistants are sorted
alphabetically by name in the selection screen. If set to “No,” the
overview is sorted by number.
Time allowed for selection (seconds)
(Repeat Program)
If a run did not proceed correctly, then the program can be
restarted. If this occurs within the time input, then all input data is
automatically re-entered, except the operator login and password.
Audio signal
If set to “Yes,” an audible signal sounds for a warning, error, and
when the program is done.
Delayed startup show as option
Set the time when the delayed startup option will be displayed
when setting up a disinfectant cycle to run.
Table 2 Left/Right Parameters
Parameter Function
Machine program (Monday through Sunday)
Select the day to automatically execute the machine cleaning
program.
Machine program (time period)
Start time for the automatic start. Either basin may have a delayed
start of up to two hours.
Water line disinfect program parameter set
Parameter set to be used when water line disinfect program has
been scheduled.