61
Loading Disinfectant
WARNING Use only RAPICIDE
®
PA High Level Disinfectant.
Check the expiration date of the disinfectant on the machine each day.
Never use disinfectant beyond the manufacturers expiration date, even if the MRC level is
acceptable.
Never use disinfectant that tests below the acceptable MRC level, even if it has not yet reached
the expiration date.
When replacing disinfectant, always wear protective clothing, gloves and eye wear. The
disinfectant manufacturer may recommend additional protection.
If leaking liquid is present and you suspect it is disinfectant, read and follow the manufacturers
instructions before attempting a cleanup.
Service: If leaking uids are present, contact a qualied service technician.
1. Press the Menu
button on the control panel.
2. Log in to the system.
3. When the Program Selection screen opens, select Replace Fluids and click the OK button.
Figure 10 Program Selection Screen
4. When the Select Fluid Type screen opens, select Disinfectant and click OK.
5. Enter the code found on the manufacturers container and press OK.
6. Remove, and discard the empty containers of Part A and Part B solutions.
7. Place the new containers of Part A and Part B solutions into the machine.