Page 31
Measure
optoCONTROL 2600
6.3.7.4 Display Scaling
The display values can be changed by the parameters Gain and Offset.
Corrected value Display value * Display gain - Display offset
You can, for example, add a constant value (offset displacement) or influence the slope
of a characteristic by a gain.
Entry occurs via Main menu > Edit program > Enter offset for display or
Enter gain for display.
i
The function Display scaling is not available in the Multi-segment measure-
ment program.
Entry of a gain should occur before any mastering or zero-setting, whereas the offset can
be modified after mastering or zero-setting.
To displace the display value add the desired displacement to the displayed displace-
ment and enter the new value at Display offset.
In addition a two-point calibration can be carried out. For the two-point calibration it is
best to use two reference pieces which correspond to the smallest and largest expected
measurements.
t
l
true measurement (set value), largest dimension
t
s
true measurement (set value), smallest dimension
d
l
Display value (actual value), largest dimension
d
s
Display value (actual value), smallest dimension
t
l
- t
s
Display gain
d
l
- d
s
Display offset
l
- Display gain * d
l
t
Example:
t
l
8.000 mm
t
s
8.005 mm
d
l
7.000 mm
d
s
7.003 mm
Display gain: 0.99800
Display offset: +0,0110 mm
The menu points Enter offset and Enter gain are not available in the
Multi-segment measurement program.
The settings Offset and Gain for the display or the analog output have no effect on the
digital value.