Measure
optoCONTROL 2600
6.6.3.3 Stop Command
Name
STOP
Description: Stops the permanent measurement output from the sensor.
Command:
Byte 1 Byte 2 Byte 3 Byte 4 hex
„+“ „+“ „+“ 0x0D 0x0D2B2B2B
„0“ „D“ „C“ „1“ 0x3143444F
0x21 0x20 0x00 0x00 0x00002021
Response:
Byte 1 Byte 2 Byte 3 Byte 4 hex Remark
„0“ „D“ „C“ „1“ 0x3143444F
0x21 0xA0 0x03 0x00 0x0003A021 Without error
0x00 0x00 0x00 0x00 0x00000000 Error code
i
Start is on when the sensor is switched on. The command Stop is volatile and is
lost when the voltage supply is switched off or the Reset command is sent.
6.6.3.4 Reset Command
Name
RESET
Description: The sensor executes a software reset. This corresponds to switch-
ing the sensor off and then on again.
Command:
Byte 1 Byte 2 Byte 3 Byte 4 hex
„+“ „+“ „+“ 0x0D 0x0D2B2B2B
„0“ „D“ „C“ „1“ 0x3143444F
0x01 0x20 0x00 0x00 0x00002001
Response:
Byte 1 Byte 2 Byte 3 Byte 4 hex Remark
„0“ „D“ „C“ „1“ 0x3143444F
0x01 0xA0 0x02 0x00 0x0002A001 Without error
0x00 0x00 0x00 0x00 0x00000000 Error code

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the MICRO-EPSILON optoCONTROL 2600 and is the answer not in the manual?

MICRO-EPSILON optoCONTROL 2600 Specifications

General IconGeneral
BrandMICRO-EPSILON
ModeloptoCONTROL 2600
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals