Measure
optoCONTROL 2600
6.6.3.5 Change the Measurement Program
Name
CHOOSE MP
Description: The sensor changes the current measurement program.
This corresponds to selecting the measurement program via the
display without the options of saving, i.e. after switching the sensor
off and on the last saved measurement program is loaded.
Command:
Byte 1 Byte 2 Byte 3 Byte 4 hex Remark
„+“ „+“ „+“ 0x0D 0x0D2B2B2B
„0“ „D“ „C“ „1“ 0x3143444F
0x23 0x20 0x01 0x00 0x00012023 Command, follow-
ing lenght
(32 bit words)
0x02 0x00 0x00 0x00 0x00000002 0 ... EDGEHL
1 ... EDGELH
2 ... DIA
3 ... GAP
4 ... SEG 2 4
5 ... MULTISEG
6 ... USER1
7 ... USER2
8 ... USER3
9 ... USER4
Response:
Byte 1 Byte 2 Byte 3 Byte 4 hex Remark
„0“ „D“ „C“ „1“ 0x3143444F
0x23 0xA0 0x03 0x00 0x0003A023 Without error
0x00 0x00 0x00 0x00 0x00000000 Error code