Measure
optoCONTROL 2600
6.6.3.8 Read out min/max Values Followed by Reset
Name
RD MINMAX RESET
Description: After the command response the min/max values are transmitted
in the range 0 ... 65519.
Then the min/max content is set to zero.
Command:
Byte 1 Byte 2 Byte 3 Byte 4 hex
„+“ „+“ „+“ 0x0D 0x0D2B2B2B
„0“ „D“ „C“ „1“ 0x3143444F
0x34 0x20 0x00 0x00 0x00002034
Response:
Byte 1 Byte 2 Byte 3 Byte 4 hex Remark
„0“ „D“ „C“ „1“ 0x3143444F
0x34 0xA0 0x04 0x00 0x0004A034 Without error
0x34 0xE0 0x03 0x00 0x0003E034 Error + 4 bytes
error code
Min [0...65519]: Binär - 0x00008B3E
Max [0...65519]: Binär - 0x00008B4B
Min/Max[mm] Min/Max[0...65519] * 40.824 / 65519 - 0.4204872