Page 52
Measure
optoCONTROL 2600
6.6.3.13 Write Measurement Program Data
Name
WR MPR TO RAM
Description: With this command measurement program data is written from
the receive buffer to the main memory. The validity of the data is
checked while this occurs. If incorrect data is found or a different
error arises, the data is not accepted into the main memory.
Command:
Byte 1 Byte 2 Byte 3 Byte 4 hex
„+“ „+“ „+“ 0x0D 0x0D2B2B2B
„0“ „D“ „C“ „1“ 0x3143444F
0x28 0x20 0x0F 0x00 0x000F2028
Description Format
Type
Bits Valid values Remark
Measurement program
number
Binary
unsigned
short
16 6 ... 9 6 - USER1
9 - USER4
Measurement program
name
ASCII
char
8x8 “A“ - “Z“, “ “, “ “,
“0“ - “9“
Only uppercase
letters. The last
space characters
are deleted. Space
characters be-
tween the letters
are replaced by “ “
(“underscore”).
Place-holder Binary
unsigned
short
16
Analog offset Binary
float
32 -10.000 ...
+10.000
Entry is made in
[ VDC ]
Analog gain Binary
float
32 -4.0000 ...
+4.0000
Display offset Binary
float
32 -99.999 ...
+99.999
Entry is made in mm
Display gain Binary
float
32 -2.000 ...
+2.000
Upper limit Binary
float
32 -168,876
...+168,876
Lower limit Binary
float
32
Upper warning Binary
float
32
Lower warning Binary
float
32
Reserve Binary
unsigned
short
16 No effect. The factory data are accept-
ed.
Measurement mode Binary
unsigned
short
16 0, 1, 2, 3, 4,
5, 6, 7
0... NORMAL
1 ... MAX CONT
2 ... MIN CONT
3 ... P-P CONT
4 ... MAX TRIG
5 ... MIN TRIG
6 ... P-P TRIG
7 ... SC1 TRIG