Version 1.10 Page 62 of 203
The available fields for the different test types are the following:
4.2.2.2 JUMP TEST
Main Parameters
Start type: Indicates if the start of the test must take place with a ‘status change’ (i.e.
entering or exiting the bars) or with an ‘external impulse’ coming from the jack sockets
(start button, signal light, etc.)
Start: indicates if at the beginning of the test the athlete must be ‘inside the area’ of the
bars or ‘outside the area’
Stop type: As for the start, the end of the test can take place because of a ‘status
change’ or an ‘external impulse’, or (in the case of time tests) the end of the test time
(e.g. a test “Jumps 15 seconds” will have the Stop type = ‘End of Time’).
Finish: for the Stop type = ‘Status change’ it is necessary to specify if, at the end of the
test, the athlete must be ‘inside the area’ of the bars or ‘outside the area’; in the other
two Stop type cases this field does not appear
Number of Jumps: appears only if the Stop type = ‘Status change’ and specifies how
many times the athlete must jump (from 1 to 99)
Test length: appears only if Stop type = ‘End of Time’ and specifies how many
minutes:seconds the test must last (from 00:01 to 09:59)
Secondary Parameters
Minimum contact time [ms]: allows to set a minimum contact time in milliseconds; if
the contact time is shorter, it is considered incorrect; if the system registers a contact
time shorter than this value, the time is added to the associated flight time (usually the
previous one). This control can be enabled/disabled by inserting the value 0.
Minimum flight time [ms]: allows to set a minimum flight time in milliseconds; if the
flight time is lower, it is considered incorrect; if the system registers a flight time shorter
than this value, the time is added to the associated contact time. This control can be
enabled/disabled by inserting the value 0.
Maximum flight time [ms]: allows to set a maximum flight time in milliseconds; if the
flight time is higher, it is considered incorrect; if the system registers a flight time longer
than this value, the time is not considered.
Starting foot: if known, the foot with which the test starts (“right” or “left”) can be
indicated, otherwise leave as “not defined”