4.2.2.6 TREADMILL GAIT TEST
Main Parameters
Start type: this can only be ‘Software command
Stop type: this can only be ‘Software command’ or ‘End of time’
Test length: appears only if Stop type = ‘End of Time’ and specifies how many
minutes:seconds the test must last (from 00:01” to 59’:59”)
Treadmill speed: indicates the speed in km/h the treadmill is set at; the setting range is
0.1-20.0 km/h with steps of 0.1.
Direction: Indicates, if the walking direction is 'Interface side' or 'Opposite side'.
Therefore, if the interface drums are at the end of the treadmill choose 'Opposite Side',
or vice versa.
Secondary Parameters
Minimum contact time [ms]: see above
Minimum flight time [ms]: see above
Maximum flight time [ms]: see above
Step length calculation: see above
Minimum step length [cm: see above
Minimum foot length [cm]: see above
Filter GaitR IN and Filter GaitR OUT [led]: see above
Automatic Data Filter: from 10% to 90% (default = deactivated); activating this
parameter, the user can check if the values for flight time, contact time, gait time, swing
time, and step length are within the average value +/- the chosen percentage. If one of
these data is out of range, the table line will be deleted from the calculation of average
values and the values on the grid will appear in strike-through format (e.g. average =
0.8, filter = 20%, all values above 0.96 and below 0.64 will be deleted).