6.1.1 2D DRIFT PROTOCOL
With its two-dimensional system (even the minimum 1 x 1 meter will suffice), the protocol can be
significantly simplified by making the patient perform just 10 jumps (5 with the right and 5 with the
left).
It is recommended to use this direction always in order to have correctly-oriented axes and
therefore a correct interpretation of the concepts of Forward/Backward and Left/Right
RX bars
RX bars
TX bars
If you notice you have not performed the test in this way, you can rotate the acquisition through
180°: view one (or both) of the tests comprising the Protocol (Right or Left Leg)
Click with the right mouse button on the bar area and select "Rotate Acquisition 180°"