MSB780(X) - User’s Guide
6.18 "intv?" Get actual the outputting interval for RUN mode
To Get actual the outputting interval for RUN mode, use the intv?<cr> command.
Example:
intv?<cr>
INTV: 1 s<cr><lf>
6.19 "avrg" Averaging time
The avrg<cr> is used to set and inspect the averaging time during which the individual mea-
surement samples are integrated to get an averaged pressure reading. The averaging time is
the total averaging time of the resulting pressure. QFE, QNH and HCP corrected pressures are
calculated from the resulting averaged pressure as well. Pressures from individual transducers
(P1, P2, P3) are not averaged.
The output reading is a running average pressure reading. The measurement is updated in nor-
mal measurement mode approximately every 1.6 seconds, depending on the avrg<cr> setting.
Example:
avrg<cr>
Averaging time = 0 s<cr><lf>
New averaging time: 100<cr>
Averaging time = 100 s<cr><lf>
wait..OK<cr><lf>
This values are averaging if is setting time 1 ... 600 s:
• P pressure.
• HCP pressure
• QFE pressure
• QNH pressure
6.20 "avrg?" Get averaging time
The avrg?<cr> is used to get setting averaging time.
Example:
43