MSB780(X) - User’s Guide
avrg?<cr>
Averaging time = 100 s<cr><lf>
6.21 “addr” Change the address in POLL mode
Addresses are required only for POLL mode. It is not supported on service port. Use the addr
a<cr> command to set barometer address, where a is address (0 . . . 255)(default = 0).
Example: the barometer is configured to address 200.
addr<cr>
Select new barometer address:200<cr>
New ADDR 200<cr><lf>
wait..OK<cr><lf>
6.22 “addr?” Get actual address in POLL mode
To Get actual address in POLL mode, use the addr?<cr> command.
Example:
addr?<cr>
ADDR: 200<cr><lf>
6.23 “open” Access to command mode in POLL mode
When all barometers on the serial RS-232 or RS-485 interface are in POLL mode the open a,
where a is address of barometer (0 . . . 255) command sets one barometer temporarity to STOP
mode so that other commands can be entered.
Example:
open 200<cr>
Command mode is opened
6.24 “close” Close command mode in POLL mode
Use the close<cr> command to switch the barometer back to the POLL mode.