MSB780(X) - User’s Guide
Table 22 continued from previous page
Problem Possible cause Solution
Display is not working,
but the barometer re-
sponds to commands
via interfaces.
Display flat cable is
not connected to the
mainboard
Open the barometer lid and reconnect the
cable.
Compensation sensor
error is signalized
Compensation sensor
is not installed
Install compensation sensor, please refer
to section 4.4 on page 25.
Compensation sensor
is faulty
If compensation sensor is installed, but
there is still the error signalized, the sen-
sor may be faulty. Replace the compensa-
tion sensor with a new one. You can order
it from MicroStep-MIS.
Other errors are sig-
nalized
Barometer is dam-
aged
Please contact MicroStep-MIS service, re-
fer to section 11.1 on page 90.
11.1 In case of problems...
It is recommended to rely on manufacturer service if you experience any problems. Even if
you have well trained personnel, it is more probable, that your barometer will be successfully
serviced in hands of the designers of the product.
When sending the product back to MicroStep-MIS please try to send a description of problems
along with the product. Pack the barometer properly, stuff the package with enough shock-
absorbing material in order to prevent it from damage during transport.
If you experience problems with the product, we would like to know about them in order to
improve the future designs.
Direct contact in case of problems with MSB780 or MSB780X: