MSB780(X) - User’s Guide
Table 16: Valid ranges for HHCP parameter
Valid ranges for HHCP parameter
Unit Minimum Maximum
m -30 30
ft -99 99
6.46 “tqfe” Set temperature for QFE corrected pressure
This command sets default temperature for calculation of the QFE corrected pressure and saves
it to FLSAH memory. The temperature is used as default after each restarting of the barome-
ter. This command is not suitable for periodic use because the FLASH has limited number of
write cycles. For setting actual temperature for QFE corrected pressure, please, use "tqfea"
command.
tqfe<cr>
TQFE:
Actual value 20.00 C<cr><lf>
Temperature scale [C, F, K]: C<cr>
Set temperature in C, ranges -80.00 .. 300.00 : 30<cr>
New temperature is: 30.00C<cr><lf>
wait..OK
Available temperature units are
◦
C,
◦
F, K.
6.47 “tqfea” Set actual temperature for QFE corrected pressure
This command sets actual temperature for calculation of the QFE corrected pressure without
saving to FLSAH memory. The temperature is used immediately. This command is suitable
especially for cases when another system sets actual temperature to MSB periodically.
6.48 “hhcp” Set height difference for HCP level
This command changes the height difference for HCP corrected pressure. The parameter may
be inputted either in ft, or in m unit. Limits for this parameter are in table 16.
Example:
hhcp<cr>
HHCP:<cr><lf>
Actual value 0.00 m<cr><lf>
Distance scale[ft, m]: m<cr>
55