MSB780(X) - User’s Guide
cset<cr>
Select sensor [ 1 2 ] : 1<cr>
CSET for sensor 1 : [ k q ] "0.1234"<cr><lf>
Set new ’k’ coefficient:1.00034<cr>
k = 1.000340<cr><lf>
Set new ’q’ coefficient:0.03<cr>
q = 0.030000<cr><lf>
wait..OK<cr><lf>
In this example, coefficients for transducer 1 were updated.
6.37 “cnf?” Get actual adjust coefficients
To read the actual values of the adjust coefficients k, q, please refer to section 6.5 on page 34.
6.38 “dpmax” Set the maximum permissible pressure difference
To set a new value of the dpmax parameter, use the dpmax<cr> command.
Example:
dpmax<cr>
DPMAX:
Actual value 1.00 hPa
Pressure scale [hPa, inHg, psi, torr, bar, mbar, mmHg, kPa, Pa, mmH2O, inH2O]: hPa<cr>
Set dpmax in hPa, ranges +0.00 .. 5.00 : 0.4<cr>
New dpmax is: 0.40 hPa
wait..OK<cr><lf>
6.39 “dpmax?” Get the maximum permissible pressure difference
To get the dpmax parameter value, use this command.
Example:
dpmax?<cr>
DPMAX: 0.40 hPa<cr><lf>
50