MSB780(X) - User’s Guide
Table 8: Range for adjust analog output
Parameter Mode / Range
1 0 .. 5 V
2 0 .. 10 V
3 0 .. 20 mA
90% set, insert value you see at output<cr><lf>
18.0058<cr>
k=0.999682 q=0.000026<cr><lf>
wait.. OK<cr><lf>
6 RS-232 and RS-485 commands
The barometer setup is done primarily via RS-232 or RS-485 (optional). Please refer to this
chapter for detailed information on commands.
6.1 Serial interface command set
Table 9: MSB780(X) commands
MSB780 or MSB780X Commands
Command Short description Page
p
poll measurement.
33
d
poll extended measurement.
33
cnf?
print configuration, probe number and firmware.
34
errors
print error status message.
37
sdi12a
set sdi-12 address.
37
help
displays help.
38
boot
restart barometer.
38
seri
set communication settings for user RS232.
38
echo1
enable echo.
38
echo0
disable echo.
39
form
change output message format.
39
form?
get actual output message format.
41
smode
change the user port start-up operating mode.
41
smode?
get actual user port start-up operating mode.
42
intv
change the outputting interval for RUN mode.
42
intv?
get actual outputting interval for RUN mode.
41
avrg
averaging time.
43
avrg?
get actual setting averaging time.
43
Continued on next page
31