Insert the anemometer with the arrow pointing upwards as shown below into the duct mounted on the right,
pushing it all the way in. Tighten the nut so that air does not escape.
Connect the supplied Ø160mm pipes. Then carefully tighten the strap so that fumes do not escape. The figure
refers to the rear side of the unit. Connect the opposite ends to the laboratory's fume extraction system.
If the laboratory fume extraction system has two 160mm inlets, use the coupling supplied
as an optional part to connect the two pipes to a 200mm outlet (refer to chapter 5.17).
Shorten the pipes to keep the exhaust duct as linear as possible. Avoid bending the pipe at
excessive angles.
The unit is not equipped with a “blowback” system. Supply it with suitable systems should
the extraction system have back pressure situations.
It is necessary to connect the unit to a fume extraction system.
The following are connected to the ventilation system: the work surface and the work
surface rear grille.