72
SAFETY SHIELD
Indicates the position of the safety shield on the work surface, which must be kept closed. If its position is
correct (glass closed), the color is green, otherwise it turns red to warn the operator and the following pop-up
with beeper appears. During the first minute the beeper rings slowly, then it becomes more frequent.
If the safety shield is closed, the message closes on its own and turns green again. Pressing “Silent” stops the
beeper and the following warning button remains on the status bar for the operator which will only
disappear when the shield is closed.
The following figure shows that, when the safety shield is open, the air speed indication also turns red (as
explained below), as it could be different than that indicated.
AIR SPEED
Indicates the instantaneous front air speed. The numerical value in the figure is only an example.
If the motor is off, the columns are not colored and no speed is displayed.
If the speed malfunctions, an alarm appears with a beeper that can be silenced. The columns turn red to warn
the operator, as in the example below.