No. Items Result
8
No miscellaneous odds and ends are allowed to be attached or affixed to the
control panel.
9
Check the power cord / power supply and all I / O ports. Ensure that all
connections are free from damage and remain clear of foreign object blockages.
There shall be no obstacles around the system and its air vent.
10
Prior to each use of a transducer, you should perform transducer cleaning and
disinfection as per the transducer operator’s manual.
11 The overall scanning environment and field must be clean.
2.2.2 Powering ON
WARNING:
If you use a transducer giving off excessive heat, it may burn the
patient.
NOTE: When you start the system or switch between transducers, you will hear clicking sounds
– this is expected behavior.
Press the power switch on the right of the system to power on the system.
Check whether the system starts normally and check the system after powering on.
No. Items Result
1 There shall no unusual sounds or smells indicating possible overheating.
2 There shall be no persistently displayed system error message.
3
There shall no evident excessive noise, discontinuous, absent or black artifacts
in the B Mode image.
4
Check if there is abnormal heat on the surface of the transducer during an
ultrasound procedure.
5
The exam date and time are the same as the system date and time, and are
displayed correctly.
CAUTION:
If the system time is not the same as the current time, misdiagnose may
occur.
2.2.3 Restarting
When any of the following abnormalities occurs with the system, the system may be able to
recover from the abnormality by power OFF/ON once again:
z An error message is displayed and does not disappear.
z The screen display is abnormal.
z The system operations are disabled.
z Restart the system to put the new software into effect.
Preparing the System 2-5

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DP-8500 and is the answer not in the manual?

Mindray DP-8500 Specifications

General IconGeneral
BrandMindray
ModelDP-8500
CategoryMedical Equipment
LanguageEnglish

Related product manuals