Appendix D Guidance and
Declaration
The system complies with the EMC standard IEC60601-1-2: 2001+A1:2004
WARNING:
The use of unapproved accessories may diminish system performance.
NOTE:
1. Use of accessories, transducers, and cables other than those specified may result
in increased emission or decreased immunity of system.
2. This system should not be used adjacent to or stacked with other equipment. If
adjacent or tacked use is necessary, this system should be observed to verify
normal operation in the configuration in which it will be used.
3. This system needs special precautions regarding EMC and needs to be installed
and put into service according to the EMC information provided below.
4. Preventing conducted RF immunity. Due to technological limitations, the conducted
RF immunity level are limited to 1Vrms level, conducted RF interference above
1Vrms may cause wrong diagnosis and measurements. We suggested that you
position this system further from sources of conducted RF noise.
5. Operation of system, in the case that the patient physiological signal is lower than
the minimum amplitude or value specified in the product specifications, may cause
inaccurate results.
6.
Portable and mobile RF communications equipment can affects this system. See tables
1, 2, 3, and 4 below.
Table 1:
GUIDANCE AND MINDRAY DECLARATION—ELECTROMAGNETIC EMISSIONS
This system is intended for use in the electromagnetic environment specified below. The
customer or the user of this system should assure that it is used in such an environment.
EMISSIONS TEST COMPLIANCE
ELECTROMAGNETIC ENVIRONMENTGUIDANCE
RF emissions
CISPR 11
Group1
This system uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
This system is suitable for use in all establishments
including domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes
Harmonic
Emissions IEC61000-
3-2
Class A
Voltage Fluctuations/
Flicker Emissions IEC
61000-3-3
Compliance
Guidance and Declaration D-1