6-2 Auxiliary Functions
6.2 Lithotrity
“Lithotrity” helps to locate the focus point of lithotrity wave during lithotrity treatment. By watching
the procedure of lithotrity in real-time and adjusting the intension and frequency of the lithotrity
wave, the harm to the patients can be reduced to the least.
To use “Lithotrity”, please enable “Lithotrity” first in “PresetÆSystem PresetÆImage Preset”.
To enter lithotrity mode: Press <Menu> to open the B Img Menu, and then click [Lithotrity] to
open the lithotrity menu and enter lithotrity mode. In M+B mode, you need to switch to B Img
Menu.
In “Lithotrity” menu:
z [Display (On/Off)]: To display or hide the lithotrity guide line.
z [Dot Size (Small/Middle/Large)]: To set the size of dots in the guide line.
z [Exit]: To exit “Lithotrity” menu.
In lithotrity mode:
z The lithotrity line is a vertical dotted line located in the middle of the screen, the position
and direction of which cannot be changed.
z The lithotrity mark is a “×” mark located on the lithotrity line which can be moved up and
down along the lithotrity line: Roll the track ball downwards to move the mark downwards,
and roll the track ball upwards to move the mark upwards.
z The depth of the mark is displayed in the lower, right corner of the screen.