Note: The Type field appears only when you have the Dual Wave
bolus or Square Wave bolus features turned on.
If you set the type to Square Wave or Dual Wave, additional settings appear.
6. If you are setting up a Square Wave bolus or Dual Wave bolus, do the
following:
For a Square Wave bolus, set the Duration of time for the bolus delivery.
For a Dual Wave bolus, adjust the Now/Square percentages as needed,
then set the Duration of time for the Square Wave portion of the bolus.
Note: If you later turn off the Dual Wave or Square Wave feature,
your existing Preset Bolus settings are still available for use.
7. Select Save.
Changing, renaming, or deleting a Preset Bolus
You cannot delete, rename, or edit a Preset Bolus while it is delivering.
Note: You cannot edit a Dual Wave or Square Wave Preset Bolus when
the Dual Wave or Square Wave features are turned off. You can, however,
rename or delete a Dual Wave or Square Wave Preset Bolus when the
features are turned off.
To change, rename, or delete a Preset Bolus:
1. Go to the Preset Bolus Setup screen.
Menu > Insulin Settings > Preset Bolus Setup
The Preset Bolus Setup screen appears, showing any existing Preset Bolus
settings.
2. Select the desired Preset Bolus.
3. Select Options.
4. Do one of the following:
bolus
Bolus 87
MP6025957-024 / A