Sensor default settings
High sensor settings
Item Default
setting
Limits Increments
High SG alert
limit
13.8 mmol/L 5.6 to 22.2 mmol/L 0.2 mmol/L
Alert before high Off - -
Alert on high Off - -
Time before high 15 minutes 5 to 30 minutes 5 minutes
Rise Alert Off - -
Rise Limit Two up
arrows
1 up arrow
(0.056 mmol/L/min)
2 up arrows
(0.111 mmol/L/min)
3 up arrows
(0.167 mmol/L/min)
•Custom limit
(0.050 to 0.275 mmol/L/min)
High Snooze 1 hour 5 minutes to 3 hours 5 minutes
Low sensor settings
Item Default
setting
Limits Increments
Low SG alert
limit
3.4 mmol/L 2.8 to 5.0 mmol/L 0.2 mmol/L
Suspend before
low
Off - -
Suspend on low Off - -
Alert before low Off - -
Alert on low Off - -
Low Snooze 20 minutes 5 minutes to 1 hour 5 minutes