Minutes from
alarm or alert
Audio Audio and vibration Vibration
0 Audio Audio and vibrate Vibrate
1 Audio Audio and vibrate Vibrate
2 Audio Audio and vibrate Vibrate
3 Audio Audio and vibrate Vibrate
4 Audio Audio and vibrate Vibrate
5 Audio Audio and vibrate Vibrate
6 Audio and vibrate Audio and vibrate Audio and vibrate
7 Audio and vibrate Audio and vibrate Audio and vibrate
8 Audio and vibrate Audio and vibrate Audio and vibrate
9 Audio and vibrate Audio and vibrate Audio and vibrate
10 Siren and vibrate Siren and vibrate Siren and vibrate
Note: The Medical device alarm sirens immediately.
Altitude range
Pump operating range is from 70.33 kPa (10.2 psiA) to 106.18 kPa (15.4 psiA)
Storage range is from 49.64 kPa (7.2 psiA) to 106.18 kPa (15.4 psiA)
Audio frequency
The following table lists the various audible tones and their corresponding
frequencies:
Tone name Frequency tolerance (+/- 1%)
Alarm 1655 Hz followed by 3310 Hz
Alternate Alarm 1850 Hz
Siren (escalated alarm) 1655 Hz, followed by 3310 Hz
Alert 934 Hz
248 Chapter 15