The following is a representative delivery accuracy curve.
0 10 20 30 40 50 60 70 80 90 100
20
15
10
5
0
5
10
15
20
25
Trumpet Curve for MMT1511/1512
Observation Window [Minutes]
Percent Error [%]
Overall Error
Maximum Error
Minimum Error
Easy Bolus
Allows user to set up and deliver a Normal Bolus when the pump is in Sleep
Mode. This is done using the and with the help of audio and vibration cues.
Audio mode range 0 to 20 increments or Max Bolus limit, whichever
comes first
Vibrate mode range 0 to 20 increments or Max Bolus limit, whichever
comes first
Default step size 0.1 unit
Adjustable step size 0.1 to 2 units per increment up to Max Bolus limit
Environmental conditions
Pump operating temperature range is from 5°C (41°F) to 40°C (104°F).
Air pressure range is from 10.2 psi to 15.4 psi (700 hPa to 1060 hPa).
product specifications and safety information
Product specifications and safety information 255