Pump screen How to enter your calibration BG
Home screen
When the calibration option is available,
you can access the Calibrate Sensor
screen. First highlight the sensor graph on
the Home screen. Then press and hold the
button to access the calibration screen.
Enter a BG meter reading specifically for
calibration.
Calibrate Sensor screen
Menu > Sensor Settings > Calibrate
Sensor
Enter a BG meter reading specifically for
calibration.
BG Meter screen
The BG Meter screen appears
automatically when your compatible Bayer
meter sends a BG meter readings to your
pump.
Select the Calibrate Sensor option to
calibrate your sensor with the current BG
meter readings.
BG screen in Event Markers
Menu > Event Markers > BG
When you enter a BG meter reading in
Event Markers, the Event Markers screen
has an option to use the BG value for
calibration.
BG field in the Bolus Wizard screen
Home screen > Bolus > Bolus Wizard
When you enter a BG meter reading to
deliver a bolus using the Bolus Wizard,
the Bolus Wizard gives you the option to
use the BG value for calibration after the
bolus is delivered.
When to calibrate
The following table describes when to calibrate your sensor.
Calibrate Description
After warm-up is
complete.
Do your first sensor calibration.
Your pump displays a Calibrate now alert within two hours
after starting a new sensor. Your first sensor glucose reading
appears about 5 to 15 minutes after you calibrate.
186 Chapter 10
MP6025957-024 / A
RELEASED