Select Edit to adjust the Bolus value and Type, if applicable. If you change
to a Square Wave bolus, you need to enter the Duration. If you change to
a Dual Wave bolus, you need to enter the Now and Square amounts, and
the Duration.
Select Rename to assign a different name to this Preset Bolus. When the
Select Name screen appears, you can select any available name from the
list.
Select Delete to delete this Preset Bolus.
Delivering a Preset Bolus
Follow these steps to deliver a Preset Bolus. You must set up Preset Bolus
deliveries before you can use the Preset Bolus feature. For more information, see
Setting up and managing Preset Bolus deliveries, on page 86.
To deliver a Preset Bolus:
1. Go to the Home screen.
2. Select Bolus.
The Bolus screen appears.
3. Select Preset Bolus.
The existing preset bolus settings appear, showing your current BG value (if
applicable) and any insulin that is still active from previous boluses. For more
information about active insulin, see About active insulin, on page 72.
4. Select the Preset Bolus you want to deliver.
5. Verify your bolus amounts, and then select Deliver Bolus.
Your pump beeps or vibrates and displays a message when your bolus starts.
Stopping a bolus delivery
The following procedures describe how to stop a Normal bolus or a Dual Wave
bolus during the Now portion delivery, and how to stop a Square Wave bolus or a
Dual Wave bolus during the Square portion delivery.
88 Chapter 4
MP6025957-024 / A
RELEASED