Menu > Utilities > Self Test
For details, see Self Test, on page 145.
6. If the Self Test does not complete successfully, or if you are concerned about
your pump, call your local representative for assistance and check your BG.
I cannot get to the Manage Settings screen
If you go to Menu > Utilities > Manage Settings, a message appears telling you
that the feature is not normally accessible and to consult your user guide. To
access the Manage Settings screen:
1. Menu > Utilities > Manage Settings
2. Simultaneously press and hold
and for about two seconds. The Manage
Settings screen appears. For more information, see Managing your pump
settings , on page 142.
My pump display times out too quickly
Your pump display times out after 15 seconds by default in order to conserve
battery power. You can increase this setting up to three minutes. Go to Menu >
Utilities > Display Options, and then adjust the Backlight setting as desired. For
more information, see Display Options, on page 141.
Note: Be aware that using a longer Backlight time causes your pump to
use more battery power. When your pump battery is low, the timeout for
the backlight on your pump screen is automatically reduced.
Where is my pump status screen?
1. To go to the Status screen, highlight and select the status bar at the top of
your Home screen.
troubleshooting
Troubleshooting 231