Fluid delivered/stroke 0.25 μL (microliter) for 0.025 unit pump stroke
0.5 μL for 0.05 unit pump stroke
2.0 μL for 0.2 unit pump stroke
Bolus Wizard feature default settings
Item Default Limits Increments
Carb units grams - -
Insulin to carb (or
exchange) ratio
None 1–200 g/u
(0.075–15.0 u/exch)
0.1 g/u for 1–9.9 g/u;
1 g/u for ratios of 10 g/u to
200 g/u
(0.001 u/exch for
0.075–0.099 u/exch
0.01 u/exch for 0.10–9.99 u/exch;
0.1 u/exch for 10–15 u/exch)
Insulin Sensitivity
Factor
None 0.3–22.2 mmol/L 0.1 mmol/L
BG Target None 3.3–13.9 mmol/L 0.1 mmol/L
Active Insulin
Time
6 hours 2 to 8 hours 15 minutes
Bolus Wizard feature specifications
There are four different formulas the Bolus Wizard feature uses to estimate a bolus,
depending on your current BG. The following formulas apply only when the carb
units are in grams.
1. If your current BG is greater than your High BG Target, the Bolus Wizard
feature subtracts active insulin from the BG correction estimate, then adds this
to the food estimate to get the total bolus estimate. However, if the result of
subtracting active insulin from BG correction estimate is a negative number
(less than zero), the total bolus estimate is based only on the food estimate.
product specifications and safety information