• On your transmitter, a green light flashes briefly and then turns off.
Note: The search process can take up to two minutes. You cannot
access your pump screens or suspend your pump during the search
process.
6. Place the transmitter next to the pump.
If the connection is successful, your pump displays a success message. If the
Sensor feature is turned on, the Connection icon
appears on the status
bar.
If your pump does not connect to your transmitter:
1. If your pump does not connect to the transmitter, do one of the following:
• Select Retry to return to the Enter Device SN screen on your pump, and
then return to step 4 of the procedure above and follow the instructions
to search again.
• Select Cancel to return to the Connect Device screen, where you can
search again using Manual Connect or Auto Connect.
2. If you have tried to connect multiple times without success, see My pump
cannot find the sensor signal, on page 234.
Deleting the transmitter from your pump
Follow this procedure to delete the transmitter from your pump. Use this process
when you are replacing your transmitter.
To delete your transmitter from your pump:
1. Go to the Manage Devices screen.
Menu > Utilities > Device Options > Manage Devices
■
setting up continuous glucose monitoring
Setting up Continuous Glucose Monitoring 181