to stop the bolus, to see details on the amount of bolus insulin delivered, or
to access the Bolus menu. The Bolus menu provides access to the Bolus
Wizard, Manual Bolus, Preset Bolus, and Insulin Settings options.
Easy Bolus
The Easy Bolus feature allows you to quickly deliver a Normal bolus using only the
button. Your pump must be in sleep mode to use the Easy Bolus feature.
Before using Easy Bolus, you must turn on the feature and set the step size. The
step size determines the number of units the bolus amount increases each time
you press the
button. Your Easy Bolus delivery is limited to 20 steps or your max
bolus limit, whichever comes first.
To help you count your Easy Bolus steps, each time you press the button, your
pump makes a different tone. There are five different tones that repeat if you use
more than five steps.
Understanding Easy Bolus step sizes
When setting up Easy Bolus, you can set the step size from 0.1 to 2.0 units. Your
step size cannot be higher than your max bolus. Set the step size to a number
that is convenient to use and easy to multiply.
The following example shows how your bolus amount is increased with each step,
or each press of the
button when using the Easy Bolus feature to deliver a
bolus. In this example, the step size is 0.5 units. For a delivery of 2.0 units, you
would need four steps, or press the button four times when using the Easy
Bolus feature.
bolus
Bolus 83