Basal patterns Maximum of 8 patterns. Each pattern covers a
24 hour period and can have up to 48 rates. Rates
are set in 30 minute increments.
Basal pattern names Fixed names: Basal 1, Basal 2, Basal 3, Basal 4, Basal 5,
Workday, Day Off, Sick Day
Increments 0.025 units per hour for basal amounts in the
range 0 to 0.975 units
0.05 units per hour for basal amounts in the range
1 to 9.95 units
0.1 units per hour for basal amounts of 10 to
35 units
BG Target
Maximum targets 8
Range 3.3 to 13.9 mmol/L
Default value for High BG
targets and Low BG targets
None
BG meter value
The most recent BG value received from the meter. If you are using a compatible
Bayer meter, this value appears on the Home screen when the Sensor feature is
off. This value also appears in the Bolus Wizard screen when setting up a bolus.
Expiration 12 minutes
Range 1.1 to 33.3 mmol/L
Bolus delivery
Bolus Speed options Standard: 1.5 units/minute
Quick: 15 units/minute
Bolus programming increments 0.025 units
•0.05 units
•0.1 units
250 Chapter 15
MP6025957-024 / A