You can set up a reminder to notify you when your next calibration is due.
For more information, see Calibration reminders, on page 133.
Setting up Auto Calibration
The Auto Calibration feature determines how to calibrate the pump when using a
wirelessly linked compatible Bayer meter. When the Auto Calibration feature is
turned on, the pump uses any BG value in the range of 2.2 to 22.2 mmol/L for
calibration. When this feature is turned off, the pump asks you every time if you
want to use the compatible Bayer meter for calibration.
To turn on Auto Calibration:
1. Go to the Auto Calibration screen.
Menu > Sensor Settings > Auto Calibration
2. Select Auto Calibration to turn on the feature.
3. Select Save.
Note: If you decide not to use the Auto Calibration in the future,
select Auto Calibration to turn off the feature.
Where to enter your calibration BG meter reading
There are several screens on the pump where you can enter a BG meter reading
for calibration. These screens are described in the following table. These options
are available only if you are using a sensor, and your transmitter is wirelessly
connected with your pump.
Note: If you are using a compatible Bayer meter and want to calibrate
using every valid BG meter reading sent to your pump, you can use the
Auto Calibration feature. For details about Auto Calibration, see Setting up
Auto Calibration, on page 185.
setting up continuous glucose monitoring
Setting up Continuous Glucose Monitoring 185
MP6025957-024 / A
RELEASED