To start a new sensor using manual connect:
1. Go to the Sensor Connections screen.
Menu > Sensor Settings > Sensor Connections
2. Select Start New Sensor.
The Start New Sensor screen appears.
3. If you have not done so already, connect the transmitter to your sensor. For
details about connecting your transmitter and sensor, see your transmitter
user guide.
Your pump searches for your transmitter signal. It can take up to six minutes
for your pump and transmitter to start communicating.
4. Select OK.
5. The Sensor warm-up started message appears on the screen. Select OK.
"Warm up..." appears on the Home screen until the sensor is ready for first
calibration. Your pump begins displaying SG data up to 15 minutes after you
successfully calibrate your sensor. For details about calibration, see Calibrating
your sensor, on page 184.
If you receive a message that your pump cannot find the sensor signal,
continue to the next section.
If your pump cannot find the sensor signal:
1. If your pump cannot find the sensor signal, follow the instructions on your
pump screen. Your pump guides you through the following steps:
a. Disconnect and reconnect the transmitter from the sensor. Pay attention
to the transmitter, and notice if the transmitter light blinks when
connected to the sensor. If the transmitter light does not blink, you need
to charge your transmitter.
b. Move your pump closer to your transmitter. It can take up to 15 minutes
for your pump to find the sensor signal.
c. If your pump is still unable to find the sensor signal, make sure you are
away from any electronic devices that might cause interference.
setting up continuous glucose monitoring
Setting up Continuous Glucose Monitoring 183
MP6025957-024 / A
RELEASED