Title and text Explanation Next steps
Pump error
Delivery stopped. Settings
unchanged. Select OK to
continue. See User Guide.
Your pump
encountered an
error but a restart is
not necessary. The
issue is resolved.
Your settings are
not changed.
Select OK to resume basal
delivery.
•If the pump was delivering a
bolus or filling the cannula, check
Daily History and reevaluate your
need for insulin.
If this alarm recurs frequently,
write down the error code
displayed on the alarm screen
(you can also find it in your
Alarm History) and contact your
local representative for assistance
with your pump.
Pump restarted
Delivery stopped. Settings
unchanged. Select OK to
continue. See User Guide.
Your pump has
encountered a
problem and has
restarted. Your
settings have not
been changed.
Select OK to continue.
•If the pump was delivering a
bolus or filling the cannula, check
Daily History and re-evaluate your
need for insulin.
If this alarm recurs frequently,
write down the error code
displayed on the alarm screen
(you can also find it in your
Alarm History) and contact your
local representative for assistance
with your pump.
Replace battery
Battery life less than 30
minutes. To ensure insulin
delivery, replace battery
now.
Battery life is low
and will be
exhausted within 30
minutes.
Select OK to clear the alert.
Replace the AA battery.
alarms, alerts, and messages
Alarms, alerts, and messages 213
MP6025957-024 / A