Title and text Explanation Next steps
Rise Alert
Sensor glucose rising
rapidly.
Your SG value has
been rising as fast
or faster than your
preset Rise Alert
Limit.
• Select OK to clear the alert.
• Monitor trend and glucose level.
• Follow instructions from your
healthcare professional.
Sensor alert occurred
Check Alarm History for
silenced alerts.
Sensor alert
occurred when Alert
Silence is on.
• Select OK to clear the alert.
• Check the Alarm History screen
to see which alerts were silenced.
For more information about
accessing the Alarm History
screen, see Alarm History, on
page 122.
• Select the alert to open the
Alarm Detail screen.
• Take action based on the
selected alert.
Sensor connected
If new sensor, select Start
New. If not, select
Reconnect.
The transmitter has
detected that you
have connected a
sensor. The pump
needs to know if
this is a new sensor
or if you have
reconnected your
old sensor.
• If you have connected a new
sensor, select Start New Sensor.
• If you have reconnected a sensor
you have been using, select
Reconnect Sensor.
• In either case, a "warm-up"
messages appears on your Home
screen, and you are prompted to
calibrate your sensor. Your pump
starts receiving your SG values
again after the two-hour
initialization is complete.
Sensor connected
Start new sensor.
The transmitter has
detected that you
have connected a
sensor.
• Select Start New Sensor.
• For more information, see
Starting the sensor, on page 182.
■
alarms, alerts, and messages
Alarms, alerts, and messages 223
MP6025957-024 / A