Recommended separation distances between the MiniMed 640G insulin
pump and common household radio transmitters
Household RF
Transmitter
Frequency Recommended
Separation
Distance (meter)
Recommended
Separation
Distance (inch)
Telephones
Cordless Household 2.4 GHz 0.3 12
Cordless Household 5.8 GHz 0.3 12
TDMA-50 Hz (cell phone) 1.9 GHz 0.3 12
TDMA-50 Hz (cell phone) 800 MHz 0.3 12
PCS (cell phone) 1.9 MHz 0.3 12
DCS (cell phone) 1.8 MHz 0.3 12
GSM (cell phone) 900 MHz 0.3 12
GSM (cell phone) 850 MHz 0.3 12
CDMA (cell phone) 800 MHz 0.3 12
Analog (cell phone) 824 MHz 0.3 12
CDMA (cell phone) 1.9 MHz 0.3 12
WiFi Networks
802.11b 2.4 GHz 1 39.5
802.11g 2.4 GHz 1 39.5
802.11n 2.4 GHz 1 39.5
Bluetooth 500 kb/s 2.4 GHz 0.1 3.93
ZigBee 250 kb/s 2.4 GHz 0.1 3.93