The Bolus Wizard screen shows your current BG meter reading (if applicable)
and any insulin that is still active from previous boluses. For more information
about active insulin, see About active insulin, on page 72. For more information,
see About your compatible Bayer meter, on page 111.
Note: Be aware that if you already have the Bolus Wizard open prior
to wirelessly sending your BG meter reading to your pump, you
must close the Bolus Wizard and open it again, in order for that
reading to appear.
2. If you are not using a wirelessly connected compatible Bayer meter, you can
select BG to manually enter your BG meter reading.
Note: If you choose not to enter a BG meter reading, three dashes
appear on the screen instead.
3. Select Carbs to enter the amount of carbs in your food.
4. Review your calculated bolus amount in the Bolus field. If you want to
change the bolus amount, select Bolus and make your desired change.
Remember, if there is a correction bolus amount calculated, you are not able
to give a Square Wave bolus.
Note: If you change your bolus amount, the word “Modified”
appears next to the new bolus amount.
5. Select Next to review your bolus information.
6. Select Square.
The Bolus Wizard screen appears with your bolus amounts.
7. To change the time period over which your bolus is delivered, select
Duration to adjust the time. The duration can be from 30 minutes to 8 hours,
in 15 minute increments.
8. Select Deliver Bolus to start your bolus.
78 Chapter 4
MP6025957-024 / A
RELEASED