EasyManua.ls Logo

NewSonic SonoDur2 - Page 10

Default Icon
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
NewSonic SonoDur2 Issue 04 09/2016 Page 10
Besides surface roughness, material properties such as texture, mechanical tensions, layer structures
and underground also play a role for measurement value variations and deviations from nominal
values.
The above information is based on experience where the real practical situation has to be tested on
the material and part in question.
The permissible limiting deviations for average values on hardness reference blocks apply in the
assessment of measurement accuracy of UCI devices - refer to table below (from DIN 50159-1/2):
Hardness
Scale
Limiting Deviation
%
< 250 HV
250 HV 500 HV
500 HV 800 HV
> 800 HV
HV 0.1
6
7
8
9
HV 0.3
6
7
8
9
HV 0.8
5
5
6
7
HV 1
5
5
6
7
HV 5
5
5
5
5
HV 10
5
5
5
5
Table 2-4
All SonoDur2 probes must meet internal standards with max. ± 3% from 3 -5 measurements on
hardness reference blocks (see Technical Data SonoDur2, Page 65).
In accordance with the UCI standard DIN 50159-1/2, hardness reference blocks with specific
dimensions are mandatory for the testing, namely a minimum thickness of 15mm and a diameter of
80mm. These blocks are often hardly obtainable. However, docking the hardness reference block to a
flat hard surface, most preferably made of steel, has much higher significance than the "correct"
dimension. Depending on the support material (wood, cloth, etc.), test force and test position, test
reference blocks can develop macroscopically very complex panel vibrations that can complicate the
performance of an UCI measurement or make it even impossible. Triangular Vickers hardness
reference blocks with a thickness of 6 mm (see image below) are very much prone to panel vibrations
and therefore it must definitely be ensured that these are always well coupled!
Figure 2.1
The best way to derive this type of influence is to observe the ranges of a measurement series.
Depending on the probe and hardness, mean values are furthermore generally more or less
significantly lower or higher than the indications provided on the reference block itself.
Caution is also needed on some Rockwell reference blocks (HRC), if they are only roughly grounded
and therefore depending on the test force, may tend to indicate too low UCI hardness values.
Attention
These triangle reference blocks need well coupling!
Your Complete Source for
Testing Equipment. Since 1969!
www.BergEng.com
Berg Engineering & Sales Company, Inc.
Tel 847-577-3980
info@bergeng.com

Table of Contents

Related product manuals