C200 SOFTWARE
Instruction Manual C200 Part2 A1.doc
optek Danulat GmbH • D-45356 Essen • Germany
page 90
Temperature Span (150R)
Set the temperature simulator to 150Ω
The display advances automatically to the next step once a stable and valid
value is recorded.
Temperature Zero (1k0)
Set the temperature simulator to 1000Ω
The display advances automatically to the next step once a stable and valid
value is recorded.
Temperature Span (1k5)
Set the temperature simulator to 1500Ω
The display advances automatically to the next step once a stable and valid
value is recorded.
Temperature Zero (2k0)
Set the temperature simulator to 2000Ω
The display advances automatically to the next step once a stable and valid
value is recorded.
Temperature Span (5k0)
Set the temperature simulator to 5000Ω
The display returns automatically to Cal. Sensor A once a stable and valid value
is recorded.
Abort Calibration
Select Yes or No
Yes selected:
– before completion of A:mV Span (+1V) frame – calibration
advances to A:T.Zero (100R) and continues.
– after completion of A:mV Span (+1V) frame – the display returns to
the Calibrate Sensor A page.
No selected – calibration continues from the point at which the
DOWNSCROLL key was pressed.
A:T.Span (5K0)
A:T.Zero (2K0)
2000
2000
Ohms
Calib
5000
5000
Ohms
Calib
A:T.Span (1K5)
1500
1500
Ohms
Calib
Cal. Sensor A
A:T.Zero (1K0)
1000
1000
Ohms
Calib
A:T.Span (150R)
Ohms
Calib
150.0
150.0
A: Abort Cal.
-----
A:T.Zero (100R)
No
Yes
CAL. SENSOR A